AMO PUERTO RICO MFG. INC. TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 08/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity: unknown/no information.The device has not returned for evaluation.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
Doctor reported that he removed an intraocular lens (iol) from a patient's right eye and replaced it with a non-johnson & johnson lens.The reason for the exchange was dysphotopsias.Reportedly, there was no capsule tear, no vitrectomy and no sutures required.No medication outside the standard of care was prescribed.No further information was provided.
 
Additional Manufacturer Narrative
Section g3, date received by manufacturer stated march 14, 2022.The correct date is march 15, 2022.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: date returned to manufacturer: jun 7, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and, no issues that could cause or contribute to visual disturbances could be identified.Based on the return condition of the lens, no further product evaluation could be performed.A product quality deficiency could not be determined.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13939767
Report Number3012236936-2022-00712
Device Sequence Number20508604
Product Code POE
UDI-Device Identifier05050474605879
UDI-Public(01)05050474605879(17)230723
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/23/2023
Device Model NumberZXT225
Device Catalogue NumberZXT225U240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer05/26/2022
07/28/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date05/26/2022
08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight80 KG
Patient RaceWhite
-
-