• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 5, 5+ 5 MM THICKNESS; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 5, 5+ 5 MM THICKNESS; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Adhesion(s) (1695); Loss of Range of Motion (2032); Scar Tissue (2060)
Date of Event 03/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2022-00913, 0001822565-2022-00915, 0001822565-2022-00916, 3007963827-2022-00080.Medical product femur cemented standard right size 5 item# 42504605802 lot# 64541285; femoral distal augment cemented size 5, 5+ 5 mm thickness item# 42556605805 lot# 64602337; femoral central cone size medium item# 42545001012 lot# 64470406; articular surface fixed bearing (cps) right 14 mm height item# 42522600414 lot# 64525555; stem extension straight splined uncemented 14 mm diameter +135 mm length item# 42560113514 item# 64542409; tibia fixed cemented right size d stem extension use required item# 42542006702 lot# 64207430; tibial augment cemented half block right medial size cd 5 mm thickness item# 42555803405 lot# 64353978; tibial augment cemented half block right lateral size cd 5 mm thickness item# 42555803205 lot# 64333007; tibial central cone size x-small item# 42545000510 lot# 64189844; stem extension 3mm offset splined uncemented 11 mm diameter +135 mm length item# 42560313511 lot# 64559287.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient developed stiffness, decreased range of motion, and scar tissue/adhesions and underwent manipulation under anesthesia approximately three months and one week post implantation.At two year follow up the patient has mild pain and stiffness with normal x-ray, no revision reported.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, b7, g3, g6, h1, h2, h3, h6, and h10 it was reported a patient had a total knee arthroplasty.Subsequently, approximately 3 months later, underwent a manipulation under anesthesia, due to stiffness.According to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthropaedics and related research (how to treat the stiff total knee arthroplasty? a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is related to limited range of motion and stiffness due to adhesions/scar tissue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 5, 5+ 5 MM THICKNESS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13942689
Report Number0001822565-2022-00914
Device Sequence Number4612209
Product Code JWH
UDI-Device Identifier00889024556911
UDI-Public(01)00889024556911(17)290731(10)64468789
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K181947
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42556805805
Device Lot Number64468789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight102 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-