INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Model Number 82-3832
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported a hakim valve was implanted in (b)(6) 2021.In mid february, the patient was not feeling good and on (b)(6) 2022 a revision surgery was done.The physician found that the joint part between the valve and the abdominal catheter was broken.Therefore, the valve was explanted and replaced.All parts were recovered.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Hakim valve was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined; however, the possible root cause could be due to a sharp or pointed object coming into contact with the device, as noted in the ifu, silicone has a low cut/tear resistance.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The hakim valve was returned for evaluation.Device history record (dhr) - lot 4088573 conformed to the specifications when released to stock.Failure analysis - the position of the cam when valve was received was 70mmh2o.The valve was visually inspected; the silicone housing at the point where the connector should be is damaged crushed/torn (possible due to a hard knock), marks were also noted on the connector.The complaint is confirmed.The root cause for the issue reported by the customer could be due to the valve receiving a hard knock damaging the silicone housing and breaking of the connector.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13942886
Report Number3013886523-2022-00122
Device Sequence Number17759366
Product Code JXG
UDI-Device Identifier10886704041825
UDI-Public(01)10886704041825(17)240831(10)4088573
Combination Product (Y/N)N
PMA/510(K) Number
K053107
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number82-3832
Device Catalogue Number823832
Device Lot Number4088573
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer06/01/2022
08/26/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date06/09/2022
09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-