| Model Number |
401743 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro mission drug-eluting stent system was selected for treatment of a calcified lesion (90 percent stenosis degree).During lesion crossing stent struts were deformed.Another stent was used to complete the intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes only the stent was returned and subjected to a technical analysis.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The stent was inflated and returned in a plastic beaker.At one side of the stent one strut is bent outwards.The delivery system was not returned for analysis.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|