Model Number 40456
Medical Device Problem Codes Failure to Advance (2524); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that guidewire tip detachment occurred.The 70% stenosed target lesion was located in the moderately tortuous and severely calcified femoral artery.A 300cm angled tip v-14 control wire was advanced to cross the lesion, however, failed to cross the lesion and the tip got detached.The detached part was removed and completed the procedure with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the complaint device was not returned by the customer, nevertheless the customer did provide pictures related to the complaint where is possible to observe a photo with the original pouch and another x-ray image that shows the distal tip section of polymer jacket detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-14 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13943721
Report Number2134265-2022-03336
Device Sequence Number17787371
Product Code DQX
UDI-Device Identifier08714729799238
UDI-Public08714729799238
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K112745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2023
Device Model Number40456
Device Catalogue Number40456
Device Lot Number0028360388
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-