• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHOCKWAVE MEDICAL, INC. SHOCKWAVE M5 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHOCKWAVE MEDICAL, INC. SHOCKWAVE M5 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER Back to Search Results
Model Number M5IVL7060
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The likely cause of the reported balloon loss of pressure and subsequent detachment of balloon or catheter component is due to patient calcium.As noted in the event description, the sharp calcium may have caused the balloon to rip and, upon withdrawal, detach.Shockwave medical has controls in place to ensure devices are built to approved procedures and meet lot release acceptance criteria prior to being distributed.A review of the manufacturing and test documentation for subject lot does not reveal any issues with the manufacturing of the device.The device passed all of shockwave medical, inc.Acceptance criteria prior to shipping.
 
Event or Problem Description
A shockwave m5 ivl catheter was used to treat an iliac artery.A balloon loss of pressure was encountered after the first cycle of pulses at 2 atmospheres.Upon removal of the ivl catheter, it was observed that a part of the balloon was detached inside of the patient.The detached portion of the balloon was pinned between the stent and the iliac to prevent movement.The physician believes that the shelf of calcium may have caught the balloon upon withdrawal causing the partial balloon detachment.To date, no patient sequelae has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKWAVE M5 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER
Common Device Name
SHOCKWAVE M5 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER
Manufacturer (Section D)
SHOCKWAVE MEDICAL, INC.
5403 betsy ross drive
santa clara CA 95054
Manufacturer (Section G)
SHOCKWAVE MEDICAL, INC.
5403 betsy ross drive
santa clara CA 95054
Manufacturer Contact
alexis weil
5403 betsy ross drive
santa clara, CA 95054
MDR Report Key13943766
Report Number3015053858-2022-00023
Device Sequence Number17220725
Product Code PPN
UDI-Device Identifier00195451000195
UDI-Public(01)00195451000195(17)230531(10)A210514D
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K180958
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberM5IVL7060
Device Catalogue NumberM5IVL7060
Device Lot NumberA210514D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.0 CHARGER BALLOON; 7.0 EVERFLEX STENT; 8.0 DORADO BALLOON
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age75 YR
Patient SexMale
-
-