BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 4.0/15; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401746
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a moderately to severely calcified lesion in the moderately tortuous distal lad.After predilatation the lesion could not be crossed with the device.In removing the orsiro, upon entering the guiding catheter, the stent dislodged and remained in the proximal lad.A dilatation balloon was inserted and inflated to crush the stent into the vessel wall.The stent delivery balloon was removed without further incident.
 
Additional Manufacturer Narrative
Combination product: yes the returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon is not well folded anymore and shows signs of inflation.Contrast medium residue was observed in the inflation- and balloon lumen up to the distal balloon shoulder, indicating application of negative pressure.Microscopic inspection showed stent imprints on the exposed balloon surface, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The stent was not returned for analysis.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The application of negative pressure to the stent system may have contributed to the dislodgement of the stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 4.0/15
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13943883
Report Number1028232-2022-01614
Device Sequence Number6701628
Product Code NIQ
UDI-Device Identifier07640130439016
UDI-Public07640130439016
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/10/2023
Device Model Number401746
Device Catalogue NumberSEE MODEL NO.
Device Lot Number05214189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-