• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 363047
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® 9nc 0.109m buffered trisodium citrate plus blood collection tubes, the device experienced a sterility issues.This event occurred three times.The following information was provided by the initial reporter.The customer stated: a part of the cap is missing for 3 tubes.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter e-mail: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 08-apr-2022.Investigation summary: bd did receive one sample and 2 photographs from the customer in support of this complaint.An evaluation of the returned sample and attached photographs was conducted and the customer's indicated failure mode of improper assembly (missing stopper) was observed.Additionally,100 retention samples from the bd inventory were visually inspected, and no defects were found.Bd was able to confirm the customer¿s indicated failure mode with the returned sample and photographs provided however, bd was not able to identify a root cause for the indicated failure mode.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.H3 other text : see h10.
 
Event or Problem Description
It was reported when using the bd vacutainer® 9nc 0.109m buffered trisodium citrate plus blood collection tubes, the device experienced a sterility issues.This event occurred three times.The following information was provided by the initial reporter.The customer stated: a part of the cap is missing for 3 tubes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13943959
Report Number9617032-2022-00255
Device Sequence Number6505317
Product Code JKA
UDI-Device Identifier50382903630474
UDI-Public50382903630474
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K013971
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2022
Device Model Number363047
Device Catalogue Number363047
Device Lot Number1326306
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-