| Model Number |
TMICL13.2 |
| Medical Device Problem Codes |
Off-Label Use (1494); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2; -14.50/1.0/088 (sphere/cylinder/axis) implantable collamer lens into the patient's eye on (b)(6) 2022.It was noticed after implantation the lens had a notch near the haptic.The lens was removed and replaced with a back up lens intraoperatively with no complications.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a vial in liquid.Visual inspection found the lens optic and haptic torn.Additional information: h6 - type of investigation 3331 - device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim#: (b)(4).
|
| |
|