• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Zonular Dehiscence (2698); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity and race: asked but not provided.Date of event: unknown as information was not provided.The best estimated date is between 1/17/2017 and 3/4/2022.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was dislocated post implantation in the left eye.Through follow-up, the customer clarified that the actual cause of subluxation was pseudoexfoliation with zonular loss.The lens moved on its own spontaneously.The surgeon explanted the lens during secondary surgery and replaced with another johnson & johnson lens (same model and diopter) with scleral fixation.1 day post-operative (op) exam showed blurred visual acuity (va) due to circulating hyphema.Vision acuity pre-op is 20/30, post op is unknown.There is no trauma to the eye.The patient received prescriptive medication; however, details not provided.The patient current status is unknown.No further information is available.A separate report is being submitted for the replacement lens reported event.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Section d9: return to manufacturer: 3/25/2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half (w/ one half missing).No issues related to or could have contributed to visual issues were observed.Based on the return condition of the lens (cut in half w/ one piece missing), no further product evaluation could be performed on the lens.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.The complaint issue reported could not be confirmed and no product deficiency could be identified.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13944760
Report Number3012236936-2022-00790
Device Sequence Number19964302
Product Code HQL
UDI-Device Identifier05050474528895
UDI-Public(01)05050474528895(17)210819
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/19/2021
Device Model NumberZA9003
Device Catalogue NumberZA90030140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight86 KG
-
-