Model Number 702-04-54E
Medical Device Problem Codes Off-Label Use (1494); Malposition of Device (2616); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
As reported: "patient dislocated and was revised.Cup was repositioned and converted to an mdm.No allegations against the implants.Original usage and revision usage attached.No further information is available." spoke to rep.The head had dislocated from the liner.The elevated rim liner and femoral head were revised to a metal mdm liner, adm/ mdm poly insert, and femoral head with sleeve.The surgeon removed 2 screws from the in-situ acetabular shell, repositioned that same shell, reimplanted the original 2 screws, and added an additional screw.Rep confirmed that no further information will be released by the hospital or surgeon.
 
Event or Problem Description
As reported: "patient dislocated and was revised.Cup was repositioned and converted to an mdm.No allegations against the implants.Original usage and revision usage attached.No further information is available." spoke to rep.The head had dislocated from the liner.The elevated rim liner and femoral head were revised to a metal mdm liner, adm/ mdm poly insert, and femoral head with sleeve.The surgeon removed 2 screws from the in-situ acetabular shell, repositioned that same shell, reimplanted the original 2 screws, and added an additional screw.Rep confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
An event regarding malposition involving a trident ii shell was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that during revision procedure, a shell was explanted and subsequently reimplanted into the same patient in a different position.The malposition event could not be confirmed.The repositioning and reimplantation of a previously implanted device is considered off label per the ifu packaged with the device at the time of manufacture: "discard all damaged or mishandled implants.Never reuse an implant, even though it may appear undamaged.It may have small defects and internal stress patterns which may lead to early failure of the device.[.] the implants are a single-use device and are provided in a sterile condition for use in an operating room by a licensed orthopaedic surgeon.[.] single use devices cannot be explanted and subsequently reimplanted as the physical forces exerted by these actions may compromise the physical integrity, dimensions and/or surface finishes of the devices.Also, sterility cannot be assured for reused devices as cleaning and re-sterilization procedures have not been verified." further information such as pre- and post-operative x-rays and the primary/revision operative reports are needed to complete the investigation for confirming the event and determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENTII TRITANIUM CLUSTER54E
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
centennial park, elstree
mahwah, NJ 07430
2018315000
MDR Report Key13945021
Report Number0002249697-2022-00459
Device Sequence Number17818051
Product Code JDI
UDI-Device Identifier07613327344073
UDI-Public07613327344073
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/13/2026
Device Model Number702-04-54E
Device Catalogue Number702-04-54E
Device Lot Number86426501A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight64 KG
-
-