• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC L3O0200 - NATURA; PROTECTOR, OSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC L3O0200 - NATURA; PROTECTOR, OSTOMY Back to Search Results
Model Number 125269
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The end user reported that the starter hole of twenty five wafers out of three boxes of wafers were off-centered.She was unable to use any of the wafers since they would not attach to pouch.An electronic mail was received from retailer and lot number was confirmed with consumer.No photo is available at this time.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions.Batch record review: lot 1f04400 was manufactured on 30/jun/2021 in mlk-3 line, with a total of (b)(4) market units (mkus).Compliance engineer performed a batch record review on 29/mar/2023, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material id 1051277 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Photograph, video and/or physical sample evaluation: no photo related to the reported problem was available for evaluation.Conclusion summary of the related event: during the event summary and impact evaluation of this nonconformance report (ncr), with the support of the triage team expertise, the most probable causes of the problem identified: method: there was no functional or dimensional test implemented to inspect the concentricity of the disc in the minilinks lines.A visual test was currently being performed, but this defect may be visually imperceptible to the operator.The appropriate actions will be taken through the corrective and preventive actions (capas).The investigation associated with related event record has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L3O0200 - NATURA
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key13948874
Report Number9618003-2022-00327
Device Sequence Number20031689
Product Code EXE
UDI-Device Identifier00768455125166
UDI-Public00768455125166
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/29/2026
Device Model Number125269
Device Lot Number1F04400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer03/31/2023
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date04/20/2023
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-