EXACTECH, INC. EQUINOXE; REVERSE 42MM GLENOSPHERE Back to Search Results
Model Number 320-01-42
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 08/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant device(s): 300-30-06 - equinoxe preserve stem 6mm; 320-42-00 - equinoxe reverse 42mm humeral liner +0; 320-10-00 - equinoxe reverse tray adapter plate tray +0; 320-15-01 - eq rev glenoid plate.
 
Event or Problem Description
As reported, approximately 2 years post op shoulder surgery, this (b)(6) male patient reported instability and 2 subluxations over the past year, 4 total since the surgery, none requiring formal reduction.The case report form indicates this event is possibly related to devices and procedure.This event report was received through clinical data collection activities.Revision has not occurred, continuing deltoid strengthening in physical therapy.Outcome is continuing.
 
Additional Manufacturer Narrative
Section h10: (h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the dislocation and subsequent closed reduction cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition and the result of loosened supporting ligaments and muscles, which allowed for dislocation occurring due to the significant ligament reconstruction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
REVERSE 42MM GLENOSPHERE
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key13949094
Report Number1038671-2022-00313
Device Sequence Number17262335
Product Code KWT
UDI-Device Identifier10885862086396
UDI-Public10885862086396
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number320-01-42
Device Catalogue Number320-01-42
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-