LUMENIS LTD. VERSAPULSE POWERSUITE 100W; HOLMIUM SURGICAL LASERS AND DELIVERY DEVICES AND ACCESSORIES Back to Search Results
Model Number VERSAPULSE POWERSUITE 100W
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the subject device dhr confirmed that the subject device was manufactured and tested according to relevant procedures, tested before release, and shipped according to manufacturer's specifications.The system was manufactured on 7-dec-2006 and installed at the customer's site on 4-jan-2007.A lumenis service engineer visited the site the day after the reported event, and found that the laser system's display voltage settings were below specification causing the display to go 'blank'.The voltage was re-adjusted to specification.The engineer completed preventative maintenance, cleaned aligned and calibrated system.Replaced water and di filters.Replaced damaged debris shield and the system was returned to the facility in working order.A review of subject device service records revealed that the subject device has had its annual pm calls on time, during which all required actions had been performed.A review of historical product complaints shows that the same malfunction of a display failure has not led to serious injury in the past.Although the device malfunction did not cause or contribute to any change in the patient's condition, it is uncertain if the user facility had to use the alternate device as intervention to prevent permanent damage.In an abundance of caution lumenis is reporting this as an adverse event.According to the gso expert based on the age of the system, "normal break and tear " could have likely played a role in the failure.Therefore, no additional corrective or preventive action is determined necessary.Lumenis is closing this complaint, but will continue to monitor this failure mode; complaint trending will continue to monitor per global complaint handling sop (doc no.(b)(4)) and per post marketing surveillance procedure (doc no.(b)(4)).
 
Event or Problem Description
A user facility reported that during a laser procedure in which a lumenis versapulse powersuite 100w laser system was being utilized, the screen went black.Unable to continue with the system, an alternate device was brought in to complete the procedure.The delay was approximately 25 minutes and the procedure completed without incident or complication.No report of patient complications was received, and no report was received alleging the device malfunction caused or contributed to any change in the patient's condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPULSE POWERSUITE 100W
Common Device Name
HOLMIUM SURGICAL LASERS AND DELIVERY DEVICES AND ACCESSORIES
Manufacturer (Section D)
LUMENIS LTD.
6 hakidma st.
po box 240
yokneam, 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD.
6 hakidma st.
po box 240
yokneam, 20692 04
IS   2069204
Manufacturer Contact
yoav wimisner
6 hakidma st.
po box 240
yokneam, 20692-04
IS   2069204
MDR Report Key13949352
Report Number3004135191-2022-00011
Device Sequence Number6581065
Product Code GEX
UDI-Device Identifier07290109140551
UDI-Public07290109140551
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERSAPULSE POWERSUITE 100W
Device Catalogue Number0638-803-01
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Initial Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-