Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2022-00888.Concomitant medical products: tibial articular surface provisional right size ef, item#: 42527000505, lot#: 63743436.The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tasp fractured.Attempts to obtain additional information have been made; however, no more information is available at this time.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Verified item lot combination and reviewed manufacturing date as returned item # 42527000515 lot # 63681980 exhibits signs of repeated use and the post feature has fractured off all pieces were returned reference attached photographs.The device history records for item # 42527000505 lot # 63743436 were reviewed for deviations and/ or anomalies with no anomalies/ deviations identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|