Model Number 121410
Medical Device Problem Codes Self-Activation or Keying (1557); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one monopty disposable core biopsy instruments was returned for evaluation.During visual evaluation, the device appeared to have blood residue and the device was received in second rotational position with the arrow in the ready window.Upon functional testing, the device was able to be fully primed with no issues and the arrow was visible in the ready window.The device was functionally tested by cocking and firing the device 18 times into a chicken breast for a total of 18 tests.The device was able to prime and fire properly.Upon priming the device for the 19th sample the device fired after the second rotation was completed.The device was disassembled, and the internal parts were inspected.Therefore, the investigation for the identified self-activation is confirmed as the device fired upon priming the device.However, the reported failure to fire is inconclusive as the device self-activated during priming the device.A definitive root cause for the alleged failure to fire and identified self-activation could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiration date: 01/2024).
 
Event or Problem Description
It was reported that during an ultrasound-guided breast biopsy through normal density tissue, the device allegedly failed to fire.It was further reported that the buttons was felt stuck.A coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOPTY
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13951805
Report Number2020394-2022-00238
Device Sequence Number17787336
Product Code KNW
UDI-Device Identifier00801741084584
UDI-Public(01)00801741084584
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number121410
Device Catalogue Number121410
Device Lot NumberREFP2759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2022
Initial Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-