• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORP NEURAGEN NERVE GUIDE 2MM ID X 3CM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORP NEURAGEN NERVE GUIDE 2MM ID X 3CM LENGTH Back to Search Results
Model Number PNG230
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Purulent Discharge (1812); Skin Infection (4544); Swelling/ Edema (4577)
Date of Event 10/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
This is 2 of 2 reports.Other mfg report no.: 1121308-2022-00019.A facility reported that a (b)(6) male patient with (png130) neuragen nerve guide had tissue post infection.It was noted that the patient had redness, swelling and purulent drainage of the hand.No culture was taken however, the infection was treated with oral antibiotics and wound care.Medical notes: "a (b)(6) male came to the clinic on (b)(6) 2021, requesting to be seen due to concerns for infection to his hand.He is 5 weeks out from zone 2 fds/fdp repair to right index, middle and ring finger.He has been working with therapy.He reports concerns for redness, swelling and purulent drainage.He occasionally notices a malodor.He was seen by the hand surgery pa and given 1 week prescription for clindamycin.Pain is actual and suspected with numeric pain score of 4.On the examination of right hand, skin surrounding surgical scar is pink/erythematous.There is open wound underlying the ring finger proximal phalanx approximately 2 x 2 cm which has a small area of necrotic tissue centrally.There is some slough surrounding this but no frank purulence.No malodor nor fluctuance.No tenderness however patient is numb distal to the zone of injury due to digital nerve lacerations.No tenderness tracking into palm.Patient gently range his digits without pain.He is stiff consistent with is postoperative status although flexor tendons do fire.Digits are pink warm and well-perfused." images: 3 views of right hand obtained with no evidence of fracture and no osteomyelitic changes.Patients medications: aspirin 81 mg oral capsule, 1 tab, po, daily.
 
Additional Manufacturer Narrative
The neuragen nerve guide was not returned however retained samples were evaluated.Dhr review: no nonconformance or capa has been issued related to this subassembly lot.The manufacturing process of fg lot 4416572 ran as expected and no assignable causes associated to the manufacturing process which could contribute and/or be related to the reported incident were identified.Failure analysis: visual inspection was performed to eight (8) retain samples of finished goods lot 4416572.No anomaly was observed: package (dispenser box), sealing area, and product appearance were in good condition.Integra's scientific affairs evaluation: "neura gen was not used at the site of infection in the palmar aspect of the ring finger, that nerve had been repaired by primary suture.Report says that two neuragen conduits were used to repair radial digital nerves of the index and middle finger (very likely these were the median digital nerves).Conclusion: conduits not involved at site of infection in the ring finger.Therefore conduits not a cause of the infection.".
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURAGEN NERVE GUIDE 2MM ID X 3CM LENGTH
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORP
1100 campus drive
1100 campus drive
princeton NJ
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORP
1100 campus drive
princeton NJ
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13952018
Report Number1121308-2022-00020
Device Sequence Number6514486
Product Code JXI
UDI-Device Identifier10381780026464
UDI-Public10381780026464
Combination Product (Y/N)N
PMA/510(K) Number
K011168
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model NumberPNG230
Device Catalogue NumberPNG230
Device Lot Number4416572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date05/13/2022
Date Device Manufactured01/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age34 YR
Patient SexMale
-
-