| Model Number |
TSA5-065 |
| Medical Device Problem Codes |
Fracture (1260); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the upper titanium bone locking screw was bent.The patient heard an audible crunching noise and during follow up x-rays showed the screw had sheared off completely.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The precice lock screw has not been returned for evaluation.The lot number has not been supplied nor could the information be obtained.This investigation was performed based on a provided x-ray.During review of the provided x-ray image, it was observed that the screw was sheared off completely as indicated in the reported description of the event.A definitive cause could not be determined due to no device return.Based on investigations of previous, similar complaints, the probable cause is excessive bending force placed on the screw as a result of weight bearing activities.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|