• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA22-80/I16-40
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 03/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft.Approximately four (4) months post initial procedure possible aortic stenosis close to the renal arteries was identified.Reintervention has been scheduled.The patient's current condition is unknown.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft.One-hundred and twelve days post initial procedure possible aortic stenosis close to the renal arteries was identified.Reintervention has been scheduled.After the initial report, reintervention was completed with the implant of an afx vela suprarenal.The aortic stenosis was resolved.The patient is reported to be doing fine.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.An examination of medical records and/or medical imaging received by endologix shows the aortic stenosis and secondary endovascular procedure are confirmed.This is consistent with the reported adverse event/incident.The most likely causation for the aortic stenosis is user related due to off-label use.The initial case was off- label due to the absence of an aortic aneurysm and the pre-existing condition of aortoiliac occlusive disease (aiod).The procedure related harms for this complaint could not be determined.The final patient status was reported being stable.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.B2: outcomes attributed to ae - updated.B5: describe event or problem - updated.G3: awareness date ¿ updated.H6: health effect - impact codes ¿ remove code 4614.H6: medical device problem codes ¿ remove 3190.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13952494
Report Number2031527-2022-00067
Device Sequence Number4524116
Product Code MIH
UDI-Device Identifier00818009015006
UDI-Public(01)00818009015006(17)240806
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/06/2024
Device Model NumberBEA22-80/I16-40
Device Lot Number2496061002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-