• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028485
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).The device has not been returned for investigation.The device history record of batch number 74l2101280 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.
 
Event or Problem Description
The problem per product concern, at least 4 plus patients have arrived from the or with dual chamber pleur-evac lot 74l2101280 have had an air leak noted.When further investigated the air leak was isolated to a system problem.Red attachment connector leaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key13953386
Report Number3004365956-2022-00028
Device Sequence Number4186427
Product Code KDQ
UDI-Device Identifier14026704631671
UDI-Public14026704631671
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2024
Device Model NumberIPN028485
Device Catalogue NumberS-1102-08LF
Device Lot Number74L2101280
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Initial Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-