Patient information: unknown/ not provided.Date of event: unknown/ not provided.If implanted, give date: unknown/ not provided.If explanted, give date: unknown/ not provided.A leaked sample cup was received.Customer provided photos were received.No complaint product device was received.A sample cup was received wrapped in bubble wrap.The bubble wrap was noticeably wet, indicating considerable leakage from the cup during shipment/prior to receipt at the site.A small amount of liquid remained inside the cup.The interior of both the cup and its lid were inspected, however no solid particles were detected.The reported foreign material could not be located for further evaluation.The customer provided photos were reviewed.The assessment revealed an image that shows a superposition of what appears to be a foreign particle colorless and with an almost rectangular shape that measures approximately 2 x 1.5 mm and is located in the mid periphery of the lens at 4:00.Unable to determine if it is located in the anterior or posterior surface of the iol.The source, nature and potential clinical impact of the observed phenomena cannot be determined from a picture assessment.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no other complaints have been received for this po.Therefore, no escalations are required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.An attempt has been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|