• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE MODULUS XLIF INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE MODULUS XLIF INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 1221050P2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vertebral Fracture (4520)
Date of Event 02/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received via advanced materials study in extreme lateral interbody fusion clinical review that a patient underwent an extreme lumbar interbody fusion at l4/5 on (b)(6) 2021.On (b)(6) 2022 during a follow up an acute fracture was visualized through the l5 body using an mri lumbar spine without contrast done.Treatment included physical therapy.
 
Additional Manufacturer Narrative
No device or radiographs were provided to confirm the complaint.The patient's post-operative physical activity and bone quality is unknown.It is unknown if the patient suffered a fall.No root cause could be determined though review of the reported information suggests implant selection, poor bone quality, or excessive post-operative physical activity as possible causes or contributors.No additional investigation can be completed.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery include: fracture of the vertebra"."warnings, cautions and precautions: correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone." "patient education the patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE MODULUS XLIF INTERBODY SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key13956045
Report Number2031966-2022-00066
Device Sequence Number4176687
Product Code MAX
UDI-Device Identifier00887517726605
UDI-Public887517726605
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K201692
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1221050P2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Initial Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
Patient SexFemale
-
-