| Model Number |
2738 |
| Health Effect - Clinical Code |
Electric Shock (2554)
|
| Date of Event |
03/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device allegedly caused an unexpected shock to the patient when the device experienced a sudden frequency jump.There was no reported patient consequence or adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the device allegedly caused an unexpected shock to the patient when the device experienced a sudden frequency jump.There was no reported patient consequence or adverse event.Djo is awaiting the return of this device.Additional reporting on this event will be provided as a supplemental report to this document if and when the device is returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
After performing functional and electrical testing, the technician was unable to duplicate the reported complaint.The product met manufacturer specifications.Possible root causes include moisture or misapplication of the electrodes.
|
| |
|