Model Number 2738
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 03/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device allegedly caused an unexpected shock to the patient when the device experienced a sudden frequency jump.There was no reported patient consequence or adverse event.
 
Additional Manufacturer Narrative
It was reported that the device allegedly caused an unexpected shock to the patient when the device experienced a sudden frequency jump.There was no reported patient consequence or adverse event.Djo is awaiting the return of this device.Additional reporting on this event will be provided as a supplemental report to this document if and when the device is returned for evaluation.
 
Additional Manufacturer Narrative
After performing functional and electrical testing, the technician was unable to duplicate the reported complaint.The product met manufacturer specifications.Possible root causes include moisture or misapplication of the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELECT TRANSPORT COMBO PKG US STD
Common Device Name
STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Manufacturer (Section D)
DJO LLC
5919 sea otter place
ste 200
carlsbad CA 92010
Manufacturer (Section G)
DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
carretera libre tijuana tecate
20230 parque industrial el flo
tijuana, b.c. 22245
MX   22245
Manufacturer Contact
jim pomeroy
5919 sea otter place
ste 200
carlsbad, CA 92010
MDR Report Key13956061
Report Number9616086-2022-00008
Device Sequence Number12002166
Product Code IMG
UDI-Device Identifier00888912021883
UDI-Public00888912021883
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K040285
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2738
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-