• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-400-25
Medical Device Problem Codes Activation Failure (3270); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the pipeline could not be opened in the blood vessel and the stent was difficult to withdraw from the catheter.The resistance was in the middle section.The catheter was flushed continuously with heparanized saline.The physician released the load (slack) in the system in an attempt to resolve the issue.The catheter nor the pushwire were damaged.The pipeline was not used off-label and the devices were prepared as indicated in the ifu.There were no patient symptoms or complications associated with the event.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm with a max diameter of 13mm and a 8mm neck diameter.It was noted the patient's vessel tortuosity was severe.A dapt was administered but the pru level was unknown.The angiographic result post procedure showed internal carotid aneurysm.Ancillary devices include: 6f 115 navien.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that the distal section did not open.
 
Additional Manufacturer Narrative
Product analysis: ¿ as found condition: the pipeline flex embolization device and marksman catheter was returned for analysis within a shipping box, within a plastic bio-pouch, and within an opened pipeline flex inner pouch (b218428).¿ visual inspection/damage location details: the pipeline flex embolization device was returned within the marksman catheter.The pi peline flex pusher was found extending out from within the marksman catheter hub for ~195.0cm.The pipeline flex pusher was found stuck at the marksman catheter hub.The marksman catheter body was found kinked at ~7.8cm from the distal tip.The marksman catheter distal tip was found in good condition.The pipeline flex embolization device could not be pushed or pulled out from within the marksman catheter.Therefore, the marksman catheter was dissected (cut) to remove the stuck pipeline flex embolization device.No bends or kinks were found with the pipeline flex pusher.The pusher was found intact.The pusher resheathing pad was found in good condition.The pipeline flex braid proximal and distal ends were found open, but damaged (frayed).The pipeline flex pusher sleeves were found damaged.No bends or kinks were found with the pipeline flex pusher tip coil.¿ testing/analysis: none ¿ conclusion: based on the device analysis and reported information, the customer¿s ¿catheter resistance¿ report was confirmed.It is possible the damage found with the marksman catheter (kinking) and the patient¿s ¿severe¿ vessel tortuosity contributed to the event.Advancing/retrieving the device against resistance can cause damage (fraying) to the pipeline flex braid.The investigation determined that this event was similar to those already investigated, and another investigation is not necessary.Regarding the customer¿s ¿failure/incomplete open distal¿ report, the issue could not be confirmed as the device has been fully deployed and re-sheathed.In addition, the returned braid was found open.It is likely the damage found with the pipeline flex braid and the patient¿s ¿severe¿ vessel tortuosity contributed to the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13956659
Report Number2029214-2022-00526
Device Sequence Number17187770
Product Code OUT
UDI-Device Identifier00847536015968
UDI-Public00847536015968
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2024
Device Model NumberPED-400-25
Device Catalogue NumberPED-400-25
Device Lot NumberB218428
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer04/02/2022
07/12/2022
Initial Report FDA Received Date03/30/2022
Supplement Report FDA Received Date04/28/2022
07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
-
-