Model Number 60005
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Site reported the light adjustable lens was explanted due to zone formation from ambient uv light exposure.Another lal was implanted during the exchange.Rxsight's first awareness of this event was (b)(6) 2022.Device history record for the lens was reviewed.No issues were noted.The explanted lens was returned for evaluation.Visual inspection and optical testing of the returned lens confirmed the presence of an ambient zone on the lens, which is indicative of exposure of the lens to ambient uv light prior to lock in.
 
Event or Problem Description
Site reported the light adjustable lens was explanted due to zone formation from ambient uv light exposure.Another lal was implanted during the exchange.Rxsight's first awareness of this event was (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
kelli carrillo
100 columbia
aliso viejo, CA 92656
MDR Report Key13956801
Report Number3012712027-2022-00013
Device Sequence Number5054063
Product Code PZK
UDI-Device Identifier00818806020180
UDI-Public01008188060201801120060217230531
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL02-001663
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-