MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938741
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a pinhole leak was observed in the lower right seam of a 3000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag.The leak was discovered in the overwrapping when getting ready to ship the bag out.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from october 11, 2021 ¿ october 12, 2021.H10: the device was received for evaluation.Unaided visual inspection was performed which observed a tear at the lower right front side edge area of the bag.Functional testing was performed which revealed a leak in the affected area.Additional magnified inspection verified a tear/hole where the leak occurred.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13956858
MDR Text Key288279097
Report Number1416980-2022-01438
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477251
UDI-Public(01)00085412477251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH938741
Device Lot Number60323485
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received03/30/2022
Supplement Dates Manufacturer Received04/15/2022
Supplement Dates FDA Received05/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-