• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-MAGNUM PF CUP 56ODX50ID; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-MAGNUM PF CUP 56ODX50ID; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problems Inflammation (1932); Pain (1994); Metal Related Pathology (4530)
Event Date 08/26/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: cat#: 139259; m2a magnum 42-50m tpr insrt +6; lot#: 694020.Cat#: 11-103215; taperloc por fmrl lat 16x152; lot#: 600530.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-00779, 0001825034-2013-04570, 0001825034-2013-04571.(this was originally reported on: 0001825034 -2013 -04569).
 
Event Description
It was reported the bilateral patient has been experiencing pain, inflammation and elevated metal ion levels.The patient had a right hip revision approximately 11.5 years after initial surgery.The head, taper and stem were removed and replaced.No further information has been provided to date.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues: increased pain, cobalt and chromium levels raised, green color fluid.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM PF CUP 56ODX50ID
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13957354
MDR Text Key288598866
Report Number0001825034-2022-00778
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberN/A
Device Catalogue NumberUS157856
Device Lot Number334820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/10/2022
Initial Date FDA Received03/30/2022
Supplement Dates Manufacturer Received04/26/2022
Supplement Dates FDA Received05/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age40 YR
Patient SexMale
-
-