MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SERUM BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SERUM BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 366668
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® serum blood collection tubes, the device experienced discolored/abnormal additive form.The following information was provided by the initial reporter.The customer stated: it was reported that tubes have water marks on them.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Initial reporter state: address information was not able to be obtained, therefore, (b)(6) was used as a place holder.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary: bd sumter received 3 customer photos for review and analysis.The photos were evaluated and the issue of additive abnormality (condensation) was not observed, the silica window shown within the pictures meets the criteria for acceptable silica window.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Therefore, bd is not able to confirm the customers reported failure mode of additive abnormality.Based on a review of batch records and the investigation results, no root cause from the manufacturing process has been identified as a contributor to additive abnormality.Based on an evaluation of severity and frequency it was determined that no corrective action is required at this time for additive abnormality.
 
Event or Problem Description
It was reported when using the bd vacutainer® serum blood collection tubes, the device experienced discolored/abnormal additive form.The following information was provided by the initial reporter.The customer stated: it was reported that tubes have water marks on them.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SERUM BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13959787
Report Number1024879-2022-00176
Device Sequence Number5053602
Product Code JKA
UDI-Device Identifier00382903666683
UDI-Public00382903666683
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2023
Device Catalogue Number366668
Device Lot Number1298002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-