SMITH & NEPHEW, INC. LEGION HK GD MOTION ISRT 11MM SZ 4-5 LT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71423352
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 03/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after tka surgery had been performed on (b)(6) 2014, the patient experienced instability on the implants.This adverse event was addressed by revision surgery on (b)(6) 2022 to explant the whole legion hk system.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device is covered in bone cement and connected to the corresponding device.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.The clinical/medical evaluation concluded that per complaint details, a hk tka revision was performed due to instability approximately 8 years post implantation.The requested clinical documentation has not been provided as of the date of this medical investigation.Per correspondence, ¿any other info was provided online, on the complaint¿.The patient¿s current health status is unknown, although it was reported that the revision was performed to ¿resolve the instability¿.The root cause of the reported event could not be further assessed.The patient impact beyond the reported laxity and subsequent revision could not be determined.No further medical assessment is warranted at this time.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.A review of the ifu for knee systems revealed in possible adverse effects that loosening, dislocation, subluxation, excessive rotation, flexion contracture, decreased range of motion, lengthening or shortening of the leg, looseness of components, unusual stress concentrations and extraneous bone can occur.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Some potential probable causes for this event could include but not limited fit/sizing, alignment, surgical technique, joint laxity, traumatic injury and procedural/user error.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION HK GD MOTION ISRT 11MM SZ 4-5 LT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13961597
Report Number1020279-2022-01422
Device Sequence Number4176325
Product Code KRO
UDI-Device Identifier00885556031230
UDI-Public00885556031230
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K081111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71423352
Device Catalogue Number71423352
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-