Model Number NCEUP37515X
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528)
Health Effect - Clinical Codes Vascular Dissection (3160); Insufficient Information (4580)
Date of Event 03/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
An attempt was made to use an nc euphora balloon catheter in a moderately tortuous, moderately calcified lesion in the proximal left main (lm) coronary artery and the left anterior descending (lad) artery.The device was inspected with no issues noted.The device passed through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force not used during delivery.It was reported that balloon deflation difficulties occurred after post dilation of a stent.The deflation difficulties occurred following the first inflation.Several techniques were made to deflate the balloon including changing the inflation device and attempts to burst the balloon with no success.With slight pressure coming off the balloon, it was possible to pull the balloon out of the stent, while partially inflated, back through the lad, through the lm and into the guide catheter.The whole system was removed together via the radial artery.Post angiogram showed a dissection in the left main.An additional sent was used to treat the dissection.No further injury was reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: lot number provided: 223719752.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: an attempt was made to use an nc euphora balloon catheter in lesion with 80% stenosis.It was not difficult to remove the protective sheath.It was not difficult to remove the packaging stylette.The lesion was pre-dilated.A 50:50 concentration of contrast / saline was used.No difficulties were noted during inflation of the device.An inflation pressure of 18 atm was applied for first inflation.The device was moved or repositioned while inflated as it would not deflate so it was pulled back into guide while still inflated.It was possible to pull the balloon out of the non mdt stent.An additional non mdt stent was used to treat the dissection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis summary: the device returned loaded in a non-medtronic hemostatic valve, guide catheter and guide extension catheter.The balloon was partially inflated on return.Contrast was visible in the balloon.Blood was visible in the guidewire lumen.Kinks were evident to the distal shaft.Proximal balloon bond was necked.The balloon passed negative prep.It was not possible to preform inflation or deflation testing due to the condition of the proximal bond.It was not possible to remove the device from the guide catheter and guide extension catheter due to the partially inflated balloon.Deformation was evident to the distal tip.No other deformation evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EUPHORA RX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13962612
Report Number9612164-2022-01246
Device Sequence Number11989636
Product Code LOX
UDI-Device Identifier00643169396272
UDI-Public00643169396272
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K141090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberNCEUP37515X
Device Catalogue NumberNCEUP37515X
Device Lot Number223719752
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer04/13/2022
05/27/2022
06/08/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date04/25/2022
06/03/2022
06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-