| Model Number |
NCEUP37515X |
| Medical Device Problem Codes |
Deflation Problem (1149); Difficult to Remove (1528)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); Insufficient Information (4580)
|
| Date of Event |
03/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An attempt was made to use an nc euphora balloon catheter in a moderately tortuous, moderately calcified lesion in the proximal left main (lm) coronary artery and the left anterior descending (lad) artery.The device was inspected with no issues noted.The device passed through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force not used during delivery.It was reported that balloon deflation difficulties occurred after post dilation of a stent.The deflation difficulties occurred following the first inflation.Several techniques were made to deflate the balloon including changing the inflation device and attempts to burst the balloon with no success.With slight pressure coming off the balloon, it was possible to pull the balloon out of the stent, while partially inflated, back through the lad, through the lm and into the guide catheter.The whole system was removed together via the radial artery.Post angiogram showed a dissection in the left main.An additional sent was used to treat the dissection.No further injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: lot number provided: 223719752.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: an attempt was made to use an nc euphora balloon catheter in lesion with 80% stenosis.It was not difficult to remove the protective sheath.It was not difficult to remove the packaging stylette.The lesion was pre-dilated.A 50:50 concentration of contrast / saline was used.No difficulties were noted during inflation of the device.An inflation pressure of 18 atm was applied for first inflation.The device was moved or repositioned while inflated as it would not deflate so it was pulled back into guide while still inflated.It was possible to pull the balloon out of the non mdt stent.An additional non mdt stent was used to treat the dissection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the device returned loaded in a non-medtronic hemostatic valve, guide catheter and guide extension catheter.The balloon was partially inflated on return.Contrast was visible in the balloon.Blood was visible in the guidewire lumen.Kinks were evident to the distal shaft.Proximal balloon bond was necked.The balloon passed negative prep.It was not possible to preform inflation or deflation testing due to the condition of the proximal bond.It was not possible to remove the device from the guide catheter and guide extension catheter due to the partially inflated balloon.Deformation was evident to the distal tip.No other deformation evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|