• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA AMISTEM H HA COATED STD STEM SIZE 6; CEMENTLESS HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA AMISTEM H HA COATED STD STEM SIZE 6; CEMENTLESS HIP STEM Back to Search Results
Model Number 01.18.136
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient came in reporting pain due to a loose stem and the cause of the loose stem is unknown.The surgeon revised the medacta stem and head with competitor components and revised the medacta liner with a medacta liner 3 years and 1 month after primary.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 03 march 2022: lot 166429: (b)(4) items manufactured and released on 06-feb-2017.Expiration date: 2022-01-25.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event since 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM H HA COATED STD STEM SIZE 6
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key13963374
Report Number3005180920-2022-00209
Device Sequence Number14648018
Product Code LZO
UDI-Device Identifier07630030804120
UDI-Public07630030804120
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K093944
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/25/2022
Device Model Number01.18.136
Device Catalogue Number01.18.136
Device Lot Number166429
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Initial Report FDA Received Date03/31/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-