| Model Number |
RE761-DRWC,LINX QUATTRO 7,GN RESOUND |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
| Date of Event |
03/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).Investigation is still in progress, a final report will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
Description of event according to initial reporter: patient symptom: redness, irritation, inflammation, blisters in the ear canal (area in contact with the domes).Time of use before symptom appeared: 3 - 4 weeks (reaction developed gradually).The skin reaction has been confirmed by doctor.Environment: according to the information provided by the dispenser, the reaction does not seem to be caused by wearing the devices in a specific environment (home, outdoor) the patient does not use any cleaning agent.The patient has not worn hearing aids before.He tried to in the past (years ago according to the shop), but desisted because he developed otitis when wearing the instruments.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).Summary of investigation findings: device history record review have been completed.No deviation or changes were found during manufacturing of the device.Clinical assessment conclusion the case has been reviewed from a clinical perspective.Customer reported: - end user is experiencing a possible allergic reaction size small surefit domes - symptoms are redness, irritation, blisters in the ear canal- where is contact with the dome.- occurred 3-4 weeks after starting using devices - medical treatment has been sought- doctor confirmed allergy -patent does not have a history of allergy, has a history of herpes zoster and has experienced allergy with a device in the past which they stopped using due to allergy - was prescribed medication- valaciclovir, diprogenta enantyum, paracetamol, and adviced to stop wearing devices -noted that doctur think that herpes zoster got worse/aggravated by use of hearing aid -hearing aids have been sent to amplifon clinical assessment of the event: clinical conclusion is that use of the dome may have aggravated a present condition, herpes zoster, however the main cause of the problem appears to be the herpes which is known to cause the symtoms described.Clinical evaluation according to (b)(4) clin eval plan&rpt,ha&tsg: allergic reaction is identified in the risk analysis and mitigated through device design by compliance with standards for biocompatibility.Still allergic reactions can occur in rare cases.The clinical evaluation does evaluate allergic reactions and the user guide includes a safety notification instructing the hearing aid wearer to contact the hcp in case of skin irritation.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusions herein are sufficient.This case is considered as a single incident and will be included in regular trending.No action haven been initiated based on this individual event.
|
| |
|
Event or Problem Description
|
|
Description of event according to initial reporter: · patient symptom: redness, irritation, inflammation, blisters in the ear canal (area in contact with the domes) · time of use before symptom appeared: 3 - 4 weeks (reaction developed gradually).· the skin reaction has been confirmed by doctor.· environment: according to the information provided by the dispenser, the reaction does not seem to be caused by wearing the devices in a specific environment (home, outdoor) · the patient does not use any cleaning agent.· the patient has not worn hearing aids before.He tried to in the past (years ago according to the shop), but desisted because he developed otitis when wearing the instruments.No additional follow up information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|