CONMED LARGO EXOSHAPE SOFT TISSUE FASTENER FEMORAL, 9 MM X 21 MM; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
1600-00-009 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the device, 1600-00-009, exoshape soft tissue fastener femoral, 9 mm x 21 mm, was being used during an unknown surgery on (b)(6) 2022 when it was reported, ¿we had another of these break today, tip broke off during implantation.¿ there was no report of injury, medical intervention, or hospitalization for the patient.Further assessment questioning found that tip broke off during implantation and fell into the surgical site, but was retrieved.Another implant was used to complete the surgery and the delay in time was "just to open another implant".This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event of implant broke during implantation was confirmed.Visual examination of the returned used device, item 1600-00-009 confirmed the reported problem, and found implant broken off and still attached to the driver shaft.Improper delivery, implant did not deploy correctly and broke off.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 1 report, regarding 1 device, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: always handle the exoshape femoral soft tissue fastener carefully.The surface of the implant must always be protected from damage during handling.Avoid contacting the implant with other tools or materials that could notch, scratch, or otherwise damage the implant surface.Damage to the implant¿s surface finish may result in loss of proper mechanical function of the device.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.
|
| |
|
Search Alerts/Recalls
|
|
|