MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMARP CV TB/IN S3 15.0; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMARP CV TB/IN S3 15.0; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 96-2033
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Insufficient Information (4580)
Date of Event 10/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Event or Problem Description
Clinical adverse event received for poly failure, tibial : wear.Event is serious and is considered mild.Event is definitely related to device and is definitely not related to procedure.Date of implantation: (b)(6) 2006.Date of event (onset): (b)(6) 2021.(left knee).Treatment: revision on 02/09/2022, with revision of tibial insert.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected:.
 
Event or Problem Description
Operative notes received indicating that "on (b)(6) 2022, the patient had a revision left total knee to address wear of articular bearing surface of internal prosthetic.The indications noted that patient had left knee pain.Depuy insert was used during the procedure.During the procedure the surgeon observed chronic foreign body synovitis with fragments of polyethylene plastic from the previous insert.The insert that was removed showed posterior lateral and some posterior medial failure including delamination pitting discoloration and fragmentation.The femoral component was well fixed.The tibial baseplate surface was intact.There was slight deformation of the patella, but no wear and it was well fixed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMARP CV TB/IN S3 15.0
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key13965737
Report Number1818910-2022-05873
Device Sequence Number14680155
Product Code NJL
UDI-Device Identifier10603295240761
UDI-Public10603295240761
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-2033
Device Catalogue Number962033
Device Lot Number8803158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer03/23/2022
04/12/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date04/07/2022
04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT POR KEEL TIB TRAY SZ2.5; PFC*SIGMA C/R NPOR FEM LT SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight66 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-