| Model Number |
Prime Plus |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The damaged board from the analyzer was returned to nova for investigation.The investigation concluded that there was damage to the power connector attached to this board.This damage is consistent with a downward "tug" on the connector.A dhr was performed on the whole prime analyzer and the board and no discrepancies were found.
|
| |
|
Event or Problem Description
|
|
Customer reported that when the prime plus analyzer was plugged in there was a burning smell.
|
| |
|
Search Alerts/Recalls
|