• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE SET; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE SET; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported foreign matter was found in the packaging of an ultrathane mac-loc locking loop multipurpose drainage set.The failure was discovered by the distributor upon receipt of the device.There was no patient involvement.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Additional information: b3- date of event is approximately 25feb2022.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Investigation evaluation: ifs miami (columbia) informed cook on 16mar2022 of an incident involving an ultrathane mac-loc locking loop multipurpose drainage set (rpn: clm-8.5-rh-npas-nt, lot: 14254422).It was reported that upon receiving the device it was noticed that there was foreign matter inside the package.This device did not make patient contact, and no adverse effects were reported to the patient due to this occurrence.Reviews of the documentation, including the complaint history, device history record, instructions for use (ifu) and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One unused and sealed device was returned for evaluation.Upon visual inspection, it was noted there was foreign matter from an unknown source near the bottom of the inner package.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for lot 14254422 records no non-conformances that could have contributed to the reported failure mode.A database search for complaints on the reported lot found one complaint from the field (mfg.Report reference #: 1820334-2022-00493) for an unrelated failure mode.Cook also reviewed product labeling.The current instructions for use [t_multi_rev5] state the following: "how supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use this product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred." evidence gathered upon review of product labeling, dmr and dhr, did not confirm that the device was manufactured out of specification or items in the lot or similar devices in the field or in house are nonconforming.Based on the information provided, examination of the returned product and the results of our investigation, it was concluded that this occurred due to a quality control deficiency.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE SET
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13966051
Report Number1820334-2022-00496
Device Sequence Number11982560
Product Code GBO
UDI-Device Identifier00827002100497
UDI-Public(01)00827002100497(17)241109(10)14254422
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/09/2024
Device Model NumberN/A
Device Catalogue NumberCLM-8.5-RH-NPAS-NT
Device Lot Number14254422
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer04/13/2022
11/11/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date04/27/2022
11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-