MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number EPS02
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Additional information was requested and the following was obtained: there is no missing piece, so no piece left in the patient.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that, during an unknown procedure, the tip of the device broke.No pieces fell into the patient.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 4/25/2022.D4 batch #: v95j85.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the eps02 device was received with the inner tube detached from the shaft sheath.In addition, the tip of the electrode was not broken as reported.Therefore, functional testing could not be performed, as the shaft could not be connected with the test handle.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key13966175
Report Number3005075853-2022-01933
Device Sequence Number4176467
Product Code GCJ
UDI-Device Identifier10705036012467
UDI-Public10705036012467
Combination Product (Y/N)N
PMA/510(K) Number
K912492
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberEPS02
Device Catalogue NumberEPS02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date04/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-