| Model Number |
3023 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim, gastrointestinal/ pelvic floor, urinary/bowel dysfunction, and urge incontinence.It was reported that the lead had high impedances and a lead replacement was performed.The caller indicated all the combinations were in the orange.Patient did want full body eligible lead, so caller indicated it was okay to replace the lead.The 3093 lead lists implant date of 2002, but the 3093 lead was not available at that time.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id 3093-28, lot# j0231031v, implanted: (b)(6) 2002, product type :lead, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received from the rep.They stated that the patient had an old 3023 lead and a 3095-10 extension.Per their conversation with tech services, this old extension requires a hex wrench which can over tighten vs.A torque wrench.The surgeon disconnected the 3023 lead from the extension and put it into the interstim x with the torque wrench.Rep called tech services afterwards and they told them the hex wrench can over tighten the lead and cause damage to the electrodes which when placed in the interstim x would result in oor impedances.The surgeon and the patient wanted to implant a surescan lead so the surgeon removed the old lead and extension and replaced it with the 978b128.The surgeon wanted to see if the ¿old¿ lead would work in case he determined that he would not be able to explant the old lead since it had been placed so long ago.New lead and battery were implanted to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id: 3093-28, lot#: j0231031v, implanted: 2002 (b)(6), product type: lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id 3093-28 lot# j0231031v serial# implanted: (b)(6) 2002.Explanted: (b)(6) 2022.Product type lead h3: analysis of the ins (s/n (b)(6) revealed that all 4 conductors output measured nominal.Shortly after the measurements were taken the device turned off due to low battery.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received from a manufacturer representative (rep).The rep reported that when the lead was connected to a replacement ins the impedances were bad.They determined that at some point a hex wrench was used instead of a torque wrench and that damaged some contacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|