MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL¿ CONNECTOR LUER LOCK C35; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL¿ CONNECTOR LUER LOCK C35; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 515200
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that prior to using a bd phaseal¿ connector luer lock c35, foreign matter was observed in the unit package.The following information was provided by the initial reporter: hair sealed into the packaging.
 
Event or Problem Description
It was reported that prior to using a bd phaseal¿ connector luer lock c35, foreign matter was observed in the unit package.The following information was provided by the initial reporter: hair sealed into the packaging.
 
Additional Manufacturer Narrative
H6: investigation summary: one photo was provided to our quality team for investigation.Upon visually inspecting the photo, a hair was observed inside the blister pack.A device history review was performed for the reported lot 2108208, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Manufacturing for this product is performed in a clean room which is kept under a positive pressure to reduce the chance of foreign matter.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.All individuals are required to follow proper gowning procedures before entering the room, including hair protection.While we could not identify a direct issue, the root cause of this incident is due to personnel not following the proper gowning procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ CONNECTOR LUER LOCK C35
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13967572
Report Number3003152976-2022-00125
Device Sequence Number4882922
Product Code ONB
UDI-Device Identifier30382905152000
UDI-Public30382905152000
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K181221
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2025
Device Model Number515200
Device Catalogue Number515200
Device Lot Number2108208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-