| Model Number |
515200 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to using a bd phaseal¿ connector luer lock c35, foreign matter was observed in the unit package.The following information was provided by the initial reporter: hair sealed into the packaging.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to using a bd phaseal¿ connector luer lock c35, foreign matter was observed in the unit package.The following information was provided by the initial reporter: hair sealed into the packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: one photo was provided to our quality team for investigation.Upon visually inspecting the photo, a hair was observed inside the blister pack.A device history review was performed for the reported lot 2108208, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Manufacturing for this product is performed in a clean room which is kept under a positive pressure to reduce the chance of foreign matter.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.All individuals are required to follow proper gowning procedures before entering the room, including hair protection.While we could not identify a direct issue, the root cause of this incident is due to personnel not following the proper gowning procedure.
|
| |
|