MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH*PROBE PLUS II SHAFT; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number EPS01
Medical Device Problem Codes Inability to Irrigate (1337); Arcing of Electrodes (2289)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that, during an unknown procedure, the device became not to energize suddenly, and became not to irrigate.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 5/4/2022.D4 batch #: u93k5t.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the eps01 device was received with the electrode tip bent and the shaft sheath damaged at the instrument tip.Due to the condition of the device returned we were unable to investigate further the irrigation issues and cautery issues as reported.A bent electrode could have caused this damage to the sheath.Caution should be taken not to retract the electrode into the sheath when it is bent.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: do not attempt to bend, sharpen, or otherwise alter the shape of the electrode tip.Doing so may cause electrode tip failure and user or patient injury.Please reference the instruction for use for more information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key13969000
Report Number3005075853-2022-01953
Device Sequence Number14696346
Product Code GCJ
UDI-Device Identifier10705036012450
UDI-Public10705036012450
Combination Product (Y/N)N
PMA/510(K) Number
K912492
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberEPS01
Device Catalogue NumberEPS01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer04/11/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date05/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-