• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-110/I20-30
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft and an afx vela suprarenal.Approximately two (2) years post initial procedure routine follow-up computerized tomography scan identified a type iiia endoleak.Reintervention was completed with the implant of an afx vela suprarenal and an afx vela infrarenal.The type iiia endoleak was successfully sealed.The patient was reported to be stable post-reintervention.
 
Event or Problem Description
Imaging received only 2 chest x-rays and a lung scan.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.The final patient status was reported as being stable post-reintervention.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13969003
Report Number2031527-2022-00070
Device Sequence Number12005792
Product Code MIH
UDI-Device Identifier00818009014542
UDI-Public(01)00818009014542(17)200624
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/24/2020
Device Model NumberBEA28-110/I20-30
Device Lot Number2264670002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date06/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL (LN 2320594016)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-