• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36705
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a bent needle occurred.The target lesion was located in the liver.A 2.0x17x15 leveen superslim was selected for use.During the procedure, the impedance rise of the radio frequency (rf) machine was completed and an abnormal needle was found after the needle was pulled out of the patient.The procedure was completed with another of the same device.There were no patient complications and the patient's status is stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the returned device matches with upn provided by the customer.One rf probe was returned for analysis (electrode).No visual damage or defects were observed on the electrode.The tines were kinked at the distal section.The device shows residue, evidence of use.The tines were extended and retracted without resistance during a functional evaluation.
 
Event or Problem Description
It was reported that a bent needle occurred.The target lesion was located in the liver.A 2.0x17x15 leveen superslim was selected for use.During the procedure, the impedance rise of the radio frequency (rf) machine was completed and an abnormal needle was found after the needle was pulled out of the patient.The procedure was completed with another of the same device.There were no patient complications and the patient's status is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEEN SUPERSLIM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13969397
Report Number2134265-2022-03608
Device Sequence Number14644480
Product Code GEI
UDI-Device Identifier08714729441694
UDI-Public08714729441694
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/11/2024
Device Model Number36705
Device Catalogue Number36705
Device Lot Number0027825706
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight78 KG
-
-