MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Migration or Expulsion of Device (1395); Material Integrity Problem (2978)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: 2021, explanted: 2022, product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 31-aug-2023, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiving morphine (5 mg/ml at 1 mg/day) via an implantable pump for unknown indications for use. It was reported that the patient's pump was flipping and the patient has intermittent return of pain. The exact date was unknown; however, the patient noted approximately 3 weeks ago. The cause of the event was undetermined. The pump pocket was revised on (b)(6) 2022. It was noted intraoperatively that the catheter on the side of pump was twisted. A 8784 was attached to existing 8780 and pump side of catheter was replaced. The catheter successfully aspirated, pump secured, and surgery successful. This issue was resolve at the time of report. The patient's weight and medical history was asked and will not be made available. The patient's status at the timeof report was alive - no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13969680
MDR Text Key290282784
Report Number3004209178-2022-04088
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured10/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/31/2022 Patient Sequence Number: 1
-
-