• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL 5.5MM HEALX ADV SP PEEK ANCHOR; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL 5.5MM HEALX ADV SP PEEK ANCHOR; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 222425
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The lot was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name, and device manufacture date were unknown.Investigation summary: the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.This complaint file was opened to document complaints derived through a journal article review.The journal article review indicated (b)(4) product failure(s).It is unknown if complaints derived from this journal article were previously reported and documented in the (b)(4) complaint system at the time of occurrence as no product code/lot number information was provided to perform the search.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported problem.Since no lot number was provided, a manufacturing record evaluation or sterile load review could not be conducted.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, (b)(4) will continue to track any related complaints within this device family as a means of monitoring the extent to which this complaint is observed in the field.
 
Event or Problem Description
It was reported by a healthcare professional through a review of a clinical research report for clinical evidence on safety and performance of 5.5mm healx adv sp peek anchor device that postoperatively to a rotator cuff repair procedure performed on an unknown date in 2019, a patient had a revision 539 days after the initial procedure that included superior capsular reconstruction of the supraspinatus capsule with tensor fascia lata autograft due to supraspinatus tear.According to the report, the patient had loosening of the graft 279 days after the initial procedure.The current status of the patient was unknown.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM HEALX ADV SP PEEK ANCHOR
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle CH-24-00
SZ   CH-2400
8472871282
MDR Report Key13970828
Report Number1221934-2022-01044
Device Sequence Number4611858
Product Code MBI
UDI-Device Identifier10886705030606
UDI-Public10886705030606
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K182941
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number222425
Device Catalogue Number222425
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer04/07/2022
Initial Report FDA Received Date03/31/2022
Supplement Report FDA Received Date04/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-