The v.A.C.® granufoam¿ dressing was not returned; therefore, a device evaluation could not be performed.Based on the information provided, it cannot be determined that the alleged bleeding event requiring surgical intervention is related to the v.A.C.® granufoam¿ dressing.The patient has a history of obesity and pilonidal cyst.Additionally, the patient had a surgical procedure in (b)(6)-2022.All end release testing of the product and packaging met specifications.Device labeling, available in print and online, states: contraindications, do not place foam dressings of the v.A.C.® therapy system directly in contact with exposed blood vessels, anastomotic sites, organs, or nerves.Warnings: bleeding: with or without using v.A.C.® therapy, certain patients are at high risk of bleeding complications.The following types of patients are at increased risk of bleeding, which, if uncontrolled, could be potentially fatal: patients who have weakened or friable blood vessels or organs in or around the wound as a result of, but not limited to: suturing of the blood vessel (native anastomosis or grafts) / organ infection.Trauma.Radiation.Patients without adequate wound hemostasis.Patients who have been administered anticoagulants or platelet aggregation inhibitors.Patients who do not have adequate tissue coverage over vascular structures.If v.A.C.® therapy is prescribed for patients who have an increased risk of bleeding complications, they should be treated and monitored in a care setting deemed appropriate by the treating physician.If active bleeding develops suddenly or in large amounts during v.A.C.® therapy, or if frank (bright red) blood is seen in the tubing or in the canister, immediately stop v.A.C.® therapy, leave dressing in place, take measures to stop the bleeding and seek immediate medical assistance.The v.A.C.® therapy units and dressings should not be used to prevent, minimize or stop vascular bleeding.Protect vessels and organs: all exposed or superficial vessels and organs in or around the wound must be completely covered and protected prior to the administration of v.A.C.® therapy.Always ensure that v.A.C.® foam dressings do not come in contact with vessels or organs.Use a thick layer of natural tissue should provide the most effective protection.If a thick layer of natural tissue is not available or is not surgically possible, multiple layers of non-adherent dressing material may be considered as an alternative, if deemed by the treating physician to provide a complete protective barrier.If using non-adherent materials, ensure they are secured in a manner that will maintain their protective position throughout therapy.Dressing changes wounds being treated with the v.A.C.® therapy system should be monitored on a regular basis.In a monitored, non-infected wound, v.A.C.® dressings should be changed every 48-72 hours, but no less than 3 times a week, with frequency adjusted by the clinician as appropriate.Infected wounds must be monitored often and very closely.For these wounds, dressings may need to be changed more often than 48-72 hours; the dressing changing intervals should be based on a continuing evaluation of the wound condition and the patient's clinical presentation, rather than a fixed schedule.
|
|
On 07-mar-2022, the following information was provided to kci by the patient: on (b)(6) 2022, after removal of the v.A.C.® granufoam¿dressing by a family member, the patient allegedly started bleeding and presented to the hospital for medical attention.The patient reported the activ.A.C.¿ ion progress¿ remote therapy monitoring system was resumed.On 10-mar-2022, the following information was provided to kci by (b)(6): on (b)(6) 2022, the patient presented to the emergency department with complaint of bleeding from his surgical site.The physician noted minor bleeding to a small subcutaneous vessel and used a suture to ligate.The activ.A.C.¿ ion progress¿ remote therapy monitoring system was resumed.On (b)(6) 2022, a device history record review for v.A.C.® granufoam¿ dressing lot number 9353894v011 was completed.All end release testing of the product and packaging met specifications.The v.A.C.® granufoam¿ dressing was not returned; therefore, a device evaluation could not be performed.
|