Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the waveguide to be broken.Functional testing found he assembled device returned a green light and produced a welcome tone.The assembled device immediately alarmed when activated.The waveguide was inspected under magnification and the origin of the crack was identified.Investigation personnel concluded that the titanium waveguide became cracked from continued use after it may have come in contact with a metallic object.Use after damage can cause the cracked waveguide to break off.No defects or damage were observed in the packaging returned that could have contributed to the reported defect.It was reported that the device received a red light and would not activate during use.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: contact between the active blade and other metal objects (hemostats, clips, staples, retractors, etc.) may result in product damage, such as a broken blade.Pieces of a broken blade may fall into the surgical cavity causing unintended tissue damage.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter, during a procedure, the dissector turned red and activation was impossible.They changed into another dissector and it worked fine.There was no patient injury.Medtronic's initial evaluation of the incident device found that the dissector had a fractured waveguide.
|