|
Model Number 10672 |
Device Problem
Break (1069)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/23/2022 |
Event Type
Death
|
Event Description
|
It was reported that shaft break occurred and the patient died.Vascular access was obtained via femoral artery.The 98% stenosed, 50x25mm, eccentric and de novo target lesion with a bend of between 45 and 90 degrees was located in the severely tortuous and moderately calcified left anterior descending artery.After a 6f non-boston scientific (bsc) guide catheter and a 0.014 non-bsc guidewire crossed the lesion, pre-dilatation was performed with a 1.5 x 15 mm and 2.0x12mm balloon catheters leaving 98% residual stenosis.A 16 x 2.25 promus elite drug-eluting stent was advanced for treatment.However, during the procedure without any force or pressure, the stent's hypotube was cut.The procedure was completed; however, after the procedure, the patient died.
|
|
Event Description
|
It was reported that shaft break occurred and the patient died.Vascular access was obtained via femoral artery.The 98% stenosed, 50x25mm, eccentric and de novo target lesion with a bend of between 45 and 90 degrees was located in the severely tortuous and moderately calcified left anterior descending artery.After a 6f non-boston scientific (bsc) guide catheter and a 0.014 non-bsc guidewire crossed the lesion, pre-dilatation was performed with a 1.5 x 15 mm and 2.0x12mm balloon catheters leaving 98% residual stenosis.A 16 x 2.25 promus elite drug-eluting stent was advanced for treatment.However, during the procedure without any force or pressure, the stent's hypotube was cut.The procedure was completed; however, after the procedure, the patient died.It was further reported that there was no resistance encountered while removing the stent.The hypotube was cut outside the patient, so the device was removed from the remaining end with no material left inside the patient.
|
|
Manufacturer Narrative
|
D6a implant date: corrected.The promus elite ous mr 16 x 2.25mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube shaft found a break situated at 20cm distal to the distal end of strain relief as well as multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that shaft break occurred and the patient died.Vascular access was obtained via femoral artery.The 98% stenosed, 50x25mm, eccentric and de novo target lesion with a bend of between 45 and 90 degrees was located in the severely tortuous and moderately calcified left anterior descending artery.After a 6f non-boston scientific (bsc) guide catheter and a 0.014 non-bsc guidewire crossed the lesion, pre-dilatation was performed with a 1.5 x 15 mm and 2.0x12mm balloon catheters leaving 98% residual stenosis.A 16 x 2.25 promus elite drug-eluting stent was advanced for treatment.However, during the procedure without any force or pressure, the stent's hypotube was cut.The procedure was completed; however, after the procedure, the patient died.It was further reported that there was no resistance encountered while removing the stent.The hypotube was cut outside the patient, so the device was removed from the remaining end with no material left inside the patient.
|
|
Search Alerts/Recalls
|
|
|