After further review of additional information received, the following sections have been updated accordingly: b4, g3, g6, h1, h2 and h6.As reported, during a medical procedure a brite tip guiding catheter (6f 0.070 jl4 100cm) was requested by the hemodynamicist at the time of uncovering it is observed that the catheter was fractured (cracked).The device was not used in the patient.Another catheter was requested for the procedure.There was no reported patient injury.The device was properly stored and opened in the sterile field.There were no complications.A non- sterile catheter ¿6f.070 jl4 100cm¿ was received for analysis.The catheter was visually inspected observing several kinked/bent conditions located approximately 29, 31 and 66 cm from the distal tip.No other damages or anomalies were noted.Inner diameter (id) and outer diameter (od) measurements were taken and were found within specification.A product history record (phr) review of lot 18069032 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The failure reported by customer as ¿brite tip/distal tip ¿ catheters ~ cracked - during prep¿ was not confirmed, the distal tip does not present any cracked condition.However, the returned unit presented with several kinked/bent conditions on the body/shaft.The exact cause of these damages could not be conclusively determined during the analysis.Shipping, storage, or handling factors may have contributed to the observed damages.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Although not intended as a mitigation of risk, information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), ¿store in a cool, dark, dry place.Do not use open or damaged packages.Remove the guiding catheter from its packaging.Inspect the guiding catheter upon removal from packaging to verify that it is undamaged.Warning: do not use a guiding catheter that has been damaged in any way.If damage is detected, replace with an undamaged guiding catheter.¿ based on the information available, product analysis, and the phr review, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
|