Model Number ZA9003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hyphema (1911); Blurred Vision (2137)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity and race: asked but not provided.If explanted; give date: n/a (not applicable).Lens remains implanted.The device was not returned for analysis as it remains implanted.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was dislocated post implantation in the left eye.Through follow-up, the customer clarified that the actual cause of subluxation was pseudoexfoliation with zonular loss.The lens moved on its own spontaneously.The surgeon explanted the lens during secondary surgery and replaced with another johnson & johnson lens (same model and higher diopter) with scleral fixation.1 day post-operative (op) exam showed blurred visual acuity (va) due to circulating hyphema.Vision acuity pre-op is 20/30, post op is unknown.There is no trauma to the eye.The patient received prescriptive medication; however, details not provided.The patient current status is unknown.No further information is available.This report captures the replacement lens, model, za9003 14.5d.A separate report is being submitted for the explanted lens, model za9003 14.0d reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13972310
Report Number3012236936-2022-00813
Device Sequence Number17750254
Product Code HQL
UDI-Device Identifier05050474528901
UDI-Public(01)05050474528901(17)260502
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2026
Device Model NumberZA9003
Device Catalogue NumberZA90030145
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Initial Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight86 KG
-
-