| Model Number |
84592 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.Two ngmc/fathom/021/bern/1ro/130 direxion fathom-16 system were selected for use.During the procedure, the wire got stuck in the catheter was noted.A new j-shape direxion catheter was opened, however, the same issue occurred.The procedure was then aborted due to this event.There were no complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The device was inspected for any damage or irregularities.The direxion had a fracture at the hub under the strain relief.The shaft was not completely separated, the inner liner was still intact.The fracture looked consistent with a bending and tensile force breaking.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.Two ngmc/fathom/021/bern/1ro/130 direxion fathom-16 system were selected for use.During the procedure, the wire got stuck in the catheter was noted.A new j-shape direxion catheter was opened, however, the same issue occurred.The procedure was then aborted due to this event.There were no complications reported.
|
| |
|